The PACT system used in the research is mainly comprised of an illumination laser, an optical parametric oscillator (OPO), a half-ring ultrasonic transducer array, data acquisition (DAQ) system and computer. For photoacoustic excitation, a 523 nm optical beam from a laser was modulated to the required wavelength by an OPO and transmitted through customized optical fiber bundles.
Characteristics
Device Name Photoacoustic Imaging System

Model

TsingPAI 定制

Characteristics

- Laser source:

  - Output wavelength: 1064nm、532nm

  - Output energy: ≥800@1064nm, ≥420mJ@532nm

  - Pulsed width: 10 ns


- Optical Parametric Oscillator (OPO)

  - Output energy: 680-950 nm @≥90mJ


- Ultrasound transducer:

  - Array type: Linear

  - Channel: 128

  - Center frequency: 5MHz ± 10%

  - Focusing: Lens focusing (50 mm in water)

  - Acquisition frame rate: 80 MSPS.